The 803PS Water Quality Sonde The 803PS Sonde is an immersible multi-electrodes measuring platform, which can accommodate up to eight different electrodes simultaneously. It fits easily into a 4-inch (100 mm) borehole and can be used to monitor water quality at depths up to 100 meters. |
![]() |
| 803PS Sonde
Electrodes The following electrodes are designed specifically to fit the 803PS Sonde Temperature sensor -5 °C to +50 °C which is a standard fitting. 830PH ...pH level 0 to 14 combination electrode. 834PH ...Redox/ORP, range ±800mV 840US ...Conductivity 0.5µS/cm to 100mS/cm. Salinity/TDS 0 to 60ppt 850OX ...Dissolved Oxygen 0 to 20ppm, 0 to 200% saturation 810TU ....Turbidity/TSS 0.2 NTU to 4000NTU 803PD ....Depth measurements 0 to 100 meter 860IS .....Ammonium ISE (Contact LMS water for ISE sensors) |
Features Up to eight different chemistry electrodes for measuring water quality can be fitted at one time. The 803PS Sonde contains its own microprocessor and control electronics, so the cable from the Sonde to the WP4007 (data logger) supplies only the power and digital interface between the instrument and the Sonde. The Sonde can withstand 300ft (100m) water depth, and can be used to measure estuaries, lakes, rivers and effluent. |
Specifications
| PH | |||
| PH | mV | ||
| Range/ Resolution | 0.00 to 14.00 pH | ± 800mV (abs) | |
| Accuracy (25°C) | ± 0.05 pH | 0.1 mV or 0.05 % rdg | |
| Response Time (98%) | 15 seconds | ||
| ORP | |||
| Range/ Resolution | ± 800mV | ||
| Accuracy (25°C) | ± 0.2mV | ||
| Response Time (98%) | 10 seconds | ||
| Conductivity/ Salinity/ TDS | |||
| Low | High | ||
| Range/ Resolution | 1.0 to 2000.00 µS/cm | 1.0 to 100.0 mS/cm | |
| Accuracy | ± 2%rdg ± 1.0 | ± 3%rdg ± 0.1 | |
| Response Time (98%) | 15 seconds | 15 Seconds | |
| Salinity/ TDS | |||
| Range/ Resolution | 0.00 to 60.00 ppt | ||
| Accuracy (25°C) | 1.50ppt | ||
| Response Time (98%) | 15 seconds | ||
| Dissolved Oxygen | |||
| Range/ Resolution | 0.00 to 20.00ppm (mg/l) 0.0% to 200.0% Saturation |
||
| Accuracy | ±0.2ppm (mg/l) ±2% |
||
| Response Time (98%) | 2 minuets | ||
| Turbidity | |||
| Low | High | ||
| Range/ Resolution | 0.3 to 100.0 NTU | 50 to 4000 NTU | |
| Accuracy | ±5%rdg ±1 NTU | ±5%rdg ±10NTU | |
| Response Time (98%) | 8 seconds | 8 Seconds | |
| TSS | |||
| Range/ Resolution | 1.3 to 250.0 mg/l (silica) | 125 to 10,000 mg/l (silica) | |
| Accuracy | ±5%rdg ±2.5mg/l | ±5%rdg ±25mg/l | |
| Response Time (98%) | 8 seconds | 8 Seconds | |
| Ammonium | |||
| Range/ Resolution | 0.1 to 200.0 ppm | ||
| Accuracy | ±10%rdg ±0.2ppm (<1ppm) ±10%rdg ±1ppm (>1ppm) |
||
| Response Time (98%) | 3 minuets | ||
| Ion Selective Electrodes (ISE) | |||
| Range/ Resolution | 0.01 to 200.00ppm |
0.01 to 200.00ppt | |
| Accuracy Sonde Electrode |
±2% rdg depends on ISE |
±2% rdg depends on ISE |
|
| Response Time (98%) | Depends on ISE type | ||
| Temperature | |||
| Sensor type | Platinum Pt100 (Din 43760) | ||
| Range/ Resolution | 0.0 to 50 °C | ||
| Accuracy | ± 0.25 °C | ||
| Response Time (98%) | 10 seconds | ||
| Depth | |||
| Electrode type | Micromachined silicon | ||
| Range/ Resolution | 0.00 to 100.00 m | ||
| Accuracy | ± 0.5 m | ||
| Power requirement | 30mA (without stirrer) 70mA (Stirrer on pulse mode) 100mA (Stirrer continously on) |
||
| Case | Acetal, Erthalyte and stainless steel 316, IP68 | ||
| Dimensions | 90mm diameter x 315mm length | ||
| Weight | 1.85Kg | ||