Conductivity/ Salinity/ TDS Measurement Measuring Conductivity All conductivity cells contain an inbuilt Pt100 temperature sensor for automatic temperature compensation. A range of both portable and in-line conductivity cells are available, depending on the ion concentration and nature of sample to be tested. Cell Types A variety of different cells are available, including beaker, dip, screw in and flow-thru. Cell Constant The cells are built to have a specific cell constant (±0.5%), but this cell constant will change if the cell becomes fouled, the following cells are available: 0.1K cell: Conductivity range (25°C): 0.1 - 100.00µS/cm 0.010/10.000mS/cm 1.0K cell: Conductivity range (25°C): 1.0 - 1000.0µS/cm 0.10 - 100.0mS/cm 10K cell: Conductivity range (25 °C): 10 - 1000.0µS/cm 1.0 - 1000mS/cm |
![]() |
| Calibration Solutions Premixed conductivity calibration solutions (±0.5%)are available: SCC1: 718µS/cm SCC3: 1408.8µS/cm SCC8: 147.0µS/cm SCC10: 111.34mS/cm |
|
Specifications
| Conductivity/ Salinity/ TDS | |||
| 0.1K | 1.0K | ||
| Electrode | 1101US | 1103US | |
| Range/ Resolution | |||
| Low | 0.1 to 200.00 µS/cm | 1.0 to 2000.0 µS/cm | |
| 5.0 to 10000 KV | 0.5 to 1000.0 KV | ||
| High | 0.050 to 10.000mS/cm | 0.50 to 100.00mS/cm | |
| 0.100 to 20.000KV | 0.010 to 2.000KV | ||
| 0 to 1000 ppm | 0.30 to 60.00 ppt (15°C) | ||
| Accuracy (25°C) | |||
| Low | ±1% rdg ± 0.1 µS | ±1% rdg ± 0.5 µS | |
| High | ±1% rdg ± 0.005 mS | ±1% rdg ± 0.05 mS | |
| ±5ppm | ±0.3ppt |
||